设为首页 |  加入收藏
首页首页 期刊简介 消息通知 编委会 电子期刊 投稿须知 广告合作 联系我们
穿戴式心电及心振信号监测系统的研制

Development of wearable ECG and SCG monitoring system

作者: 鲁亚磊  王瑞  胡敏露  王卫东 
单位:<span style="font-family:宋体">上海大学通信与信息工程学院(上海 200444)</span><p><span style="font-family:宋体">中国人民解放军总医院医学保障部(北京 100853)</span></p>
关键词: 可穿戴技术;  心脏监测系统;  心振图;  心电图;  电-机械耦联 
分类号:<span style="font-size: 14px; font-family: arial, helvetica, sans-serif;">R318.04</span>
出版年·卷·期(页码):2018·37·4(381-386)
摘要:

目的 心脏的电活动及机械活动对于研究心脏的生理功能和病理改变具有重要作用, 为此本研究研制了一套心电 (electrocardiogram, ECG) 及心振 (seismocardiogram, SCG) 信号同步监测系统。方法 采用单导联方式进行心电检测, 通过测量胸廓皮肤表面的三轴线加速度和三轴角速度来获取心振信号, ARM Cotex-M3内核的低功耗微处理器STM32作为主控芯片将数据存储至SD, 或通过低功耗蓝牙将采集的数据发送至手机端上位机并进行数据处理及波形显示。结果 对人体进行测量以获取ECGSCG信号, 结果表明系统能够实时采集、传输、显示具有较高信噪比的心电和心振信号, 并能实现信号的长期离线记录。结论 该系统具有便携、操作简单的特点, 能够长期稳定工作, 为下一步分析心脏的电活动与机械活动之间的耦联关系奠定了基础。

Objective The electrical properties and mechanical activities of the heart play an important role in the physiological and pathological cardiac research.Therefore, we develop a synchronous monitoring system for electrocardiogram (ECG) and seismocardiogram (SCG) .Methods The detection of ECG is designed by single lead and SCG is obtained by measuring the tri-axial acceleration and the tri-axial angular velocity of the skin surface at chest.A low power consumption microprocessor named STM32 with ARM Cortex-M3 kernel is used to store data to the SD card, or send data to the mobile phone by low-power bluetooth, process data and display waveform.Results ECG and SCG signals of human body are acquired, and the results show that the system can collect, transmit and display the signals of ECG and SCG with high signal-to-noise ratio in real time, and the system can complete the record of signals over a long period of time.Conclusions The monitoring system is portable and easy to operation, and can work steadily for a long time.This work makes the foundation for further analysis of the coupling relationship between the electrical activities and the mechanical activities of the heart.

参考文献:

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com