用于植入式心室辅助装置的流量拟合算法

Flow estimation algorithm for implantable ventricular assist device

作者: 王永远  陈琛  马云善 
单位:苏州大学机电学院(苏州 215000) 苏州同心医疗器械有限公司(苏州 215000)
关键词: 流量拟合;  黏度;  心衰;  人工心脏;  CH-VAD; 
分类号:R318.01
出版年·卷·期(页码):2019·38·2(177-181)
摘要:

目的 通过实验采集参数并对实验数据进行分析,得到一种流量拟合算法,在第3代全磁悬浮人工心脏CH-VAD上实现应用。方法 在实验环境设置为定常流、37℃,液体是甘油水溶液,其黏度为(3.5±0.1)cP (1 cP=10-3Pa·s),转速从2 000 r/min到4 000 r/min,流量从0.3 L/min到8.6 L/min,寻找输出流量Q(L/min)与电机电流I(mA)、转速ω(r/min)的关系。结果 在各个转速下调节阻尼,I和Q具有线性回归性,相关度R2>0.99,即I=AQ+B,其A、B与转速ω(r/min)相关。结论 算法的拟合系数具有很强的稳定性和一致性,在一定范围内算法的拟合精度不受黏度影响,拟合误差在±0.2 L/min,可以用于CH-VAD。

Objective  A flow estimation algorithm was obtained by collecting experimental parameters and analyzing experimental data, and was applied to the CH-VAD, which was the third-generation all-maglev artificial heart independently developed and produced by CH BIOMEDICAL,INC. Methods  In order to find the relationship between output flow Q(L/min ) with motor current I (mA) and speedω(r/min ) , the experimental environment was steady flow with 37 ℃.Liquid was glycerol-water solution and the viscosity was (3.5±0.1) cP (1 cP = 10-3 Pa·s).The speed of CH-VAD was from 2 000 r / min to 4 000 r / min with flow rate from 0.3 L / min to 8.6 L / min. Results  The motor current I ( mA) and output flow Q ( L / min) had linear regression under the condition of adjusting the damping at each speed, and the correlation degree R2 > 0.99, I = AQ+B, in which A and B were related to rotation speed ω ( r / min). Conclusions  Within a certain range, the accuracy of flow estimation is not affected by the viscosity, and the error of flow estimation is 0.2 L / min. The flow estimation algorithm can be used for CH-VAD

参考文献:

[1]Bui AL, Horwich TB, Fonarow GC. Epidemiology and risk profile of heartfailure[J].Nature Reviews Cardiology,2011,8(1):30-41.

[2]陈伟伟,高润霖,刘力生,等.《 中国心血管病报告 2016》概要[J].中国循环杂志,2017,32(6):521-530.

[3]王芳群,吴振海,荆腾,等. 基于左心室辅助装置和心血管耦 合模型的反流研究[J].生物医学工程学杂志,2017,34(5):752-759.

Wang FQ, Wu ZH, JingT, et al. Study on regurgitation using the coupling model of left ventricular assist device and cardiovascular system[J].Journal of Biomedical Engineering,2017,34(5):752-759.

[4]Kadakia S, Moore R, Ambur V, et al. Current status of the implantable LVAD[J].General Thoracic and Cardiovascular Surgery,2016,64(9):501-508.

[5]张岩,孙寒松,胡盛寿. 左心室辅助血泵及其临床应用研究进展[J].中国胸心血管外科临床杂志,2017,24(2):152-155.

Zhang Y, Sun HS, Hu SS. Research progress and clinical application of ventricular assist blood pump[J].Chinese Journal of Clinical Thoracic and Cardiovascular Surgery,2017,24(2):152-155.

[6]Hubbert L, Baranowski J, Delshad B, et al. Leftatrial pressure monitoring with an implantable wireless pressure sensor following implantation of a left ventricular assist device[J].ASAIO Journal,2016,63(5):e60-e65.

[7]Cotter CJ. Flow estimation using hall-effect sensors: US Patent Application,15,873,427[P].2018-5-24.

[8]Reyes C, Voskoboynikov N, Chorpenning K, et al. Accuracy of the HVADpump flow estimation algorithm[J].ASAIO Journal, 2016,62(1):15-19.

[9]Yoshizawa M , Sato T, Tanaka A , et al. Sensorless estimation of pressure head and flow of a continuous flow artificial heart based on input power and rotational speed[J].ASAIO Journal,2002,48(4):443-448.

[10]Pennings KA, Martina JR, Rodermans BF, et al. Pump flow estimation from pressure head and power uptake for the HeartAssist5, HeartMate II, and HeartWare VADs[J].ASAIO Journal,2013,59(4):420-426.

[11]李岚, 曾培, 茹伟民, 等. 基于电机参数的叶轮式人工心脏输出流量的无创性测量[J].北京生物医学工程,2001,20(2):146-148.

Li L, Zeng P, Ru WM, et al. Noninvasive pump flow estimation of an impeller artificial heart based on the motor parameters[J].Beijing Biomedical Engineering,2001,20(2):146-148.

[12]封志刚, 曾培, 茹伟民,等. 人工心脏输出流量和压力的神经网络估算法[J].中国生物医学工程学报,2002,21(6):568-572.

Feng ZG, Zeng P, Ru WM, et al. A method for computing output flow rate and pressure of artificial hearts by neural networks[J].Chinese Journal of Biomedical Engineering, 2002,21(6):568-572.

[13]李岚, 曾培, 茹伟民,等. 神经网络在叶轮式人工心脏输出流量检测中的应用[J].生物医学工程与临床,2001,5(1):12-15.

Li L, Zeng P, Ru WM, et al. The application of neural network technique for output estimation of an impeller artificial heart[J].Biomedical Engineering and Clinical Science,2001,5(1):12-15.

[14]陈智慧, 张娅, 常宇. 心室辅助下切应力对红细胞形态的影响[J].北京生物医学工程,2015,34(2):125-132.

Chen ZH, Zhang Y, Chang Y. Effect of shear stress on the changes in erythrocyte morphology under continuous flow ventricular assist device[J].Beijing Biomedical Engineering,2015,34(2):125-132.

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com