Objective A comparative study of Real Time Cell Analysis(RTCA)method with MTT method was conducted to demonstrate the feasibility of RTCA method on cytotoxicity tests of medical devices. Methods First,RTCA cytotoxicity test method based on L929 cell was established by screening cell seeding density. Then, different concentrations of DMSO diluents, extracts of standard reference materials, 56 batches of medical device products in the real world were used, from simple to complex, to compare for the consistency of RTCA and MTT methods in cytotoxicity test. The results of negative and positive controls from multiple experiments were collected to compare the repeatability of the two methods. Results The results of RTCA method showed no dose-effect relationship in the test of DMSO diluents. In the test of standard control material extracts, the results of RTCA method were completely consistent with MTT method and highly correlated (R=0.9898). In the test of 56 batches of medical device products, the consistency rate was 80.4% and only moderately correlated (R=0.6183). The repeatability of RTCA method is no better than that of MTT method. Conclusions RTCA method can effectively detect the cytotoxicity of medical devices, but some products are not applicable because their dissolves may interfere with the detection results. The applicability of RTCA as a new method to detect the cytotoxicity of medical devices needs further study.
|
[1] ISO 10993-5: 2009 Biological evaluation of medical devices—Part 5: Tests for in vitro cytotoxicity[S].Switzerland: ISO copyright office, 2009. [2] 全国医疗器械生物学评价标准化技术委员会. GB/T 16886.5—2017 医疗器械生物学评价 第5部分:体外细胞毒性试验[S]. 北京:中国标准出版社,2017. [3] Bruinink A, Luginbuehl R. Evaluation of biocompatibility using in vitro methods: interpretation and limitations[J]. Advances in Biochemical Engineering Biotechnology, 2012,126: 117–152. [4] Atienz FA, Gerets H, Tilmant K, et al. Evaluation of impedance-based label-Free technology as a tool for pharmacology and toxicology investigations[J]. Biosensors, 2013, 3(1): 132-156. [5] Limame R, Wouters A, Pauwels B, et al. Comparative analysis of dynamic cell viability, migration and invasion assessments by novel real-time technology and classic endpoint assays[J]. PLoS One, 2012, 7(10): e46536. [6] Garcia SN, Gutierrez L, McNulty A. Real-time cellular analysis as a novel approach for in vitro cytotoxicity testing of medical device extracts[J]. Journal of Biomedical Materials Research Part A, 2013, 101(7): 2097–2106. [7] 王海涛, 徐玉茵, 周静, 等.RTCA 法用于医疗器械体外细胞毒性试验的研究[J]. 毒理学杂志, 2016, 30(3): 224-226. [8] Meindl C, Absenger M, Roblegg E, et al. Suitability of cell-based label-free detection for cytotoxicity Screening of carbon nanotubes[J]. BioMed Research International, 2013,2013: 564804. [9] 段链,顾雯,王秦,等. 实时动态监测纳米氧化铜作用于CHL细胞的毒性效应[J].癌变畸变突变,2016,28 (3): 205-208, 213. Duan L, Gu W, Wang Q, et al. Real-time dynamic monitoring of cytotoxicity in nano-copper oxide-treated CHL cells[J]. Carcino genesis, Teratogenesis & Mutagenesis, 2016,28 (3): 205-208, 213. [10] 全国医疗器械生物学评价标准化技术委员会. GB/T 16886.1—2011 医疗器械生物学评价 第1部分: 风险管理过程中的评价与试验[S].北京:中国标准出版社,2011. [11] 全国医疗器械生物学评价标准化技术委员会. GB/T 16886.12—2017 医疗器械生物学评价 第12部分:样品制备与参照材料[S].北京:中国标准出版社,2017. [12] 贺学英,王蕊,王会如,等. 医疗器械细胞毒性试验方法标准化探讨[J]. 医疗装备, 2015, 28(7):9-10. [13] 高静贤,王莎莎,金梦,等. 医用超声耦合剂体外细胞毒性检测的探讨[J].中国医疗器械杂志. 2013,37(3): 210-212. Gao JX, Wang SS, Jin M, et al. In vitro cytotoxicity test of medical ultrasonic couplant[J]. Chinese Journal of Medical Instrumentation, 2013, 37(3): 210-212. [14] 宾蕾,邓慧敏,李正超, 等. RTCA与MTT法用于检测肠道微生物细胞毒性的比较研究[J]. 军事医学,2019,43(6):442-447. Bin L, Deng HM, Li ZC, et al. Detection of cytotoxicity of gut microbiota with RTCA and MTT method[J]. Military Medical Sciences,2019,43(6): 442-447. [15] 侯雪芹,张晓晶,王奇,等. 体外抗AD 药物筛选中MTT、Edu 和RTCA 法的比较研究[J]. 时珍国医国药, 2015,26(3):741-743. [16] 胡煜雯, 高静贤. GB/T 16886.5-2017中MTT细胞毒性试验方法解读[J]. 中国医疗器械信息,2018,24(21):11-12, 49. Hu YW, Gao JX. Interpretation of MTT cytotoxicity test methods in GB/T 16886.5-2017[J]. China Medical Device Information, 2018,24(21):11-12, 49.
|