设为首页 |  加入收藏
首页首页 期刊简介 消息通知 编委会 电子期刊 投稿须知 广告合作 联系我们
CT 质控中水模校准在设备使用中的保障作用

Guarantee function of water model calibration on equipment use in CT quality control

作者: 冉梓垠  王予生 
单位:北京积水潭医院设备科(北京 100035) <p>通信作者:王予生。E-mail:wangyusheng666@sina.com</p> <p>&nbsp;</p>
关键词: CT;质控;水模校准;保障作用  
分类号:R318.6
出版年·卷·期(页码):2022·41·3(315-317)
摘要:

目的  探讨水模校准在CT质控中的作用,以期为CT设备使用的保障提出建设性意见。 方法 选择3台CT设备,依据GB 17589—2011通过水模校准对设备进行质控,记录校准前后的CT值,并进行统计分析。结果 3台CT设备校准前的CT值分别为0.9±1.1、0.9±0.8、0.9±1.2,校准后的CT值分别为0.7±0.2、0.6±0.3、0.6±0.3,差异具有统计学意义(P<0.01)。结论 严格执行质控,定期进行水模校准,可以对设备使用起到保障作用。

 

Objective To discusse the role of water model calibration in CT quality control,and to provide constructive suggestions for the use of CT equipment. Methods Three sets of CT equipment were selected. According to GB 17589—2011, the quality control of the equipment was carried out through water model calibration every month, and the CT values before and after calibration were recorded and statistically analyzed. Results  The CT values of the three CT devices before calibration were 0.9±1.1, 0.9±0.8 and 0.9±1.2, respectively, and the CT values after calibration were 0.7±0.2, 0.6±0.3 and 0.6±0.3, respectively. The difference between before and after calibration was statistically significant (P < 0.01). Conclusions Strict quality control and regular water model calibration can guarantee the use of equipment.

参考文献:

[1]? 陈磊,姚庚. 规范CT设备日常养护保障临床应用效能[J].中国仪器仪表,2020(10):41-48.

Chen L,Yao G. Standardize CT routine maintenance methods to guarantee clinic application efficiency[J]. China Instrumentation,2020(10):41-48.

[2]? Anim-Sampong S,Antwi WK, Botwe B,et al.Comparison of 640-slice Aquilon ONE CT scanner's measured dosimetric parameters with ICRP dose reference levels for head, chest and abdominal CT examinations[J].Safety in Health,2016,2(1):132-133

?[3] 庄晓璇,刘鸿翔,袁田,等. CT应用质量检测结果与分析[J].中国医学装备,2019,16(10): 97–99.

Zhuang XX,Liu HX,Yuan T,et?al.Testing?results?and?analysis?of?application?quality?of?CT[J].China?Medical?Equipment,2019,16(10):97-99.

[4]? 伍健,耿建华. CT设备质量控制的现状与展望[J].中国医学装备,2018,15(11):163-166.

Wu?J,?Geng ?JH.The?current?status?and?outlook?of?quality?control?of?CT?equipment[J].China?Medical?Equipment,2018,15(11):163-166.

?[5] 张海成. CT设备的质量控制分析与应用[J].中国医学装备,2016,15(10):24-28,29

Zhang HC. Quality control analysis and application of CT equipment[J]. China?Medical?Equipment,2016(10):24-28,29.

[6]? 王树涛. CT图像的质量控制与设备维护保养[J].名医,2019(10):125.

?Wang ST. Quality control of CT image and equipment maintenance [J].Renowned Doctor,2019(10):125.

?

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com