西德后装机施源器改造的理论研究 _

Theoretical Research About the Modification of the Applicators of Remote Afterloading System Made In West Germany

作者: 秦林林; 
单位:中日友好临床医学研究所;
关键词: 遥控后装治疗机;  阴道容器;  剂量率; 
分类号:
出版年·卷·期(页码):1994·1·4(237-242)
摘要:

80年代国内引进的联邦德国布克勒公司生产的后装机,治疗宫颈癌应用三个直管型曼彻斯特容器。三枚等强度Cs ̄137源,直径4.5mm、长20mm在离容器顶端10mm处水平横卧,临床应用中常引起阴道等部位的并发症。我们将阴道平行型容器改造成直立型,剂量分布更为合理。作者从理论上分析了直立型容器的优越性。以Ir(192)和Cs(137)源为例,对5cm官腔A点剂量率后者分别提高19.4%和37.5%;对6cm宫腔增加量分别为26.2%和54.4%。在宫颈口下1~2cm范围内阴道壁平均剂量直立型约为平行型的60%左右,大大降低了阴道等组织烧伤的危险,临床应用得到满意的疗效。

Since Buchler remote afterloading system made in Germany was introduced into our country in the eighties,three parallel applicators have been used in treatment of cervix cancer. Three C sources,4. 5mm diameter and 20mm length, with equal intensity, were placed horizontally 10mm away from the top of the applicators. Complications of vaginal injury often happened when clinical used. we modified the applicators from parallel to erective, so that the dose distribution was more reasonable. In this paper the anal...

参考文献:

中国期刊全文数据库 共找到 1 条

[1] 何传泰,要守静,沈建华. 应用Buchler高剂量率后装机直立型阴道器治疗宫颈癌82例初步报告[J]. 中华放射肿瘤学杂志, 1990,(03) .
 