设为首页 |  加入收藏
首页首页 期刊简介 消息通知 编委会 电子期刊 投稿须知 广告合作 联系我们
血管介入式全身热疗计划软件的研制

Development of Software for Treatment Planning in Interventional Whole Body Hyperthermia

作者:               贾得巍  刘静          
单位:           清华大学医学院生物医学工程系(北京100084)    
关键词:           血管介入式全身热疗;治疗计划软件;房室模型;生物传热;COM组件;混合编程      
分类号:
出版年·卷·期(页码):2010·29·6(613-619)
摘要:

血管介入式全身热疗有望在晚期和扩散型肿瘤的治疗中发挥重要作用,相应临床手术的高效实施需要发展治疗计划软件来指导加热剂量的精确给定。本文从已经过验证的人体全身生物传热房室模型出发,开发出一套基于C#.Net与Matlab COM组件的计算机软件。在Matlab中完成热量守恒和差分离散,实现人体全身热效应的数值计算和后处理功能;在C#.Net中通过Heath Carter体型及成分评估策略、IO文件操作、Solidworks丰富了前处理自定义操作、可视化效果后处理和报告功能。本软件对全身热疗的基本问题提出了系统化的解决方案并将其分解为前处理、计算、后处理等几大模块并建立了评估的基本流程,计算案例成功地将全身热疗计划评估通过友好软件用户接口进行了标准化。

Interventional whole body hyperthermia (IWBH) is a very promising way for the treatment of late stage and metastatic cancer.A highly efficiency IWBH operation requires to develop a treatment planning software in order to guide a precise delivery of heating dosage.Based on the well demonstrated compartment bioheat transfer model for the whole human body,this article successfully develops a software by way of C#.Net and Matlab COM technique.The Numerical analysis of energy conservation and finite difference,as well as post processing modules was established in Matlab.In C#.Net,the applications of Heath Carter somatype analysis,IO file manipulation,and Solidworks aided realized the visualization,post-processing and reporting function.The present system provided a systemic solution to the fundamental issues involved in WBH by classifying the task into several main categories such as preprocessing,analysis and postprocessing.The calculation cases further illustrated the standardized work flow for the treatment planning procedures.

参考文献:

[1]Chen XL,Wang F,Bao YQ.An accidental burn resulted from whole body hyperthermia therapy[J].Burns,2007,33(2):252-254.

[2]Xiang SH,Chen BW,He XH,et al.Development of interventional whole body hyperthermia system//Part II:In vivo animal experimental evaluation[J].Forschung im Ingenieurwesen,2009,73(3):139-146.
[3]Jay O,Reardon FD,Webb P,et al.Estimating changes in mean body temperature for humans during exercise using core and skin temperatures is inaccurate even with a correction factor[J].J Appl Physiol,2007,103(2):443-451.
[4]Xiang SH,Chen BW,Zhou YX,et al.Development of interventional whole body hyperthermia system//Part I:Device fabrication and performance evaluation[J].Forschung im Ingenieurwesen,2008,72(3):145-152.
[5]Kistemaker JA,Den Hartog EA and Daanen HAM.Reliability of an infrared forehead skin thermometer for core temperature measurements[J].Journal of Medical Engineering & Technology,2006,30(4):252-261.
[6]吕永钢,刘静,闰井夫.红外热图监测微波辐照式全身热疗中体表温度的变化[J].中国组织工程研究与临床康复,2007,11(35):7019-7023.
[7]Liu J,Zhou YX,Deng ZS.Sinusoidal heating method to noninvasively measure tissue perfusion[J].IEEE Trans Biomed Eng,2002,49(8):867-877.
[8]Ashvinikumar VM,Brent EE,Patricia LR,et al.A phantom tissue system for the calibration of perfusion measurements[J].ASME Journal of Biomechanical Engineering,2008,130(5):051002.
[9] Carter JEL,Heath BH.Somatotyping-development and applications[M].Cambridge:Cambridge University Press,1990:454.
[10]Ronco AL,Mendoza B,Varas X,et al.Somatotype and risk of breast cancer:a case-control study in uruguay[J].Revista Brasileira de Epidemiologia,2008,11:215-227.
[11]Singh SP.Somatotype and disease-A review[J].The Anthropologist:International Journal of Contemporary and Applied Studies of Man,2007:251–261.
[12]Shier D,Butler J,Lewis R.Hole’s Human Anatomy & Physiology[M].12th ed.Dubuque:McGraw-Hill,2010:1120.
[13]Neimark MA,Konstas AA,Laine AF,et al.Integration of jugular venous return and circle of Willis in a theoretical human model of selective brain cooling[J].J Appl Physiol,2007,103(5):1837-1847.
[14]Kakuta N,Yokoyama S,Mabuchi K.Human thermal models for evaluating infrared images[J].Engineering in Medicine and Biology Magazine,IEEE,2002,21(6):65-72.

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com