2450MHz水冷式微波消融天线比吸收率的实验研究

Experimental Research on Specific Absorption Rate of Water-Cooled Microwave Antenna with 2450MHz in Ablation

作者: 艾海明  吴水才  杨春兰  高宏建  赵磊  曾毅 
单位:北京工业大学生命学院生物医学工程中心(北京100124)
关键词: microwave  ablation;  specific  absorption  rate;  temperature  field  simulation 
分类号:
出版年·卷·期(页码):2011·30·1(39-44)
摘要:

探讨2450MHz微波天线比吸收率(SAR)的空间分布规律,以掌握微波天线的加热能力。以离体猪肝作为微波热消融实验对象,分别采用5组常用微波功率40W、50W、60W、70W和80W进行加热,每组以40例样本做热消融实验。实验中,每例样本被精确固定布放16个测温针并且分别进行前向和后向测温。结果表明,SAR前、后向分布具有非对称性,从天线中心向四周有规律地衰减,并且SAR与加热功率之间没有线性关系,但数值随着加热功率增大而增大。近场点的温度模拟结果与实测值具有较好的一致性,从而验证了SAR检测结果的准确性。因此,该实验研究微波天线SAR分布的方法是有效的,SAR检测结果可用于温度场计算机模拟中的热载荷。

In order to understand the heating ability of microwave antenna, the spatial distribution pattern of special absorption rate was explored with 2450MHz microwave antenna. Five-group ablated experiments with the heating power of 40W,50W,60W,70W and 80W were performed to in-vitro pig livers. Forty samples of thermal ablation were repeated for each group. In each test, sixteen thermometers were precisely located and the temperature measurements were performed both forward and backward directions. Experimental results demonstrated that the SAR were asymmetric in forward and backward directions and decreased regularly around from antenna center. There was no linear relationship between SAR and heating power however the value of SAR increased with the heating power. Based on good consistency between temperature simulation and the real temperature measurement, in the field near temperature center, the accuracy of the SAR examination was evaluated. In conclusion, experimental study of SAR distribution of microwave antenna is effective and detection results of SAR can be applied to thermal load in computer simulation of temperature field.

参考文献:

[1]Qun Nan, Xiaoyong Yang, Youjun Liu, et al. Numerical simulation on microwave ablation with a water-cooled antenna[C].Wuhan: ICBBE,2007:698-701.
[2]南群,夏雅琴,刘有军,等. 微波消融治疗肝癌中的若干热学问题[J]. 北京工业大学学报,2008,34(5):551-555.
[3]梁萍,董宝玮,于晓玲, 等. 超声引导下植入式微波热场在体模中的研究[J]. 中华超声影像学杂志,2002,11(4):236-239.
[4]任新颖. 肿瘤热疗中组织温度场测量方法及关键技术研究[D].  北京:北京工业大学,2008.
[5]Cundin LX, Roach WP, N Millenbaugh. Empirical comparison of pennes’ bio-heat equation[C].  Washington:Proceedings of SPIE,2009,7175:71751601-71751609.
[6]Sherar MD, Gladman AS, Davidson SR, et al. Infrared thermographic SAR measurements of interstitial hyperthermia applicators: errors due to thermal conduction and convection[J].  Int J Hyperthermia,2004,20(5):539-555.
[7]Liang P, Dong B, Yu X, et al. Computer-aided dynamic simulation of microwave-induced thermal distribution in coagulation of liver cancer[J]. IEEE Transactions on Biomedical Engineering,2001,48(7):821-829.
[8]Puccini S, Bar NK, Bublat M, et al.  Simulations of thermal tissue coagulation and their value for the planning and monitoring of laser-induced interstitial thermotherapy[J]. Magnetic Resonance in Medicine,2003,49:351-362.
 

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com