如何预防X线管损坏

How to Prevent the X-Ray Tube from Damage

作者: 裴作升    刘海洋    杨义耀 
单位:襄樊职业技术学院医学院(湖北襄樊441021)
关键词: X线管;阴极损坏;阳极损坏;预防方法 
分类号:
出版年·卷·期(页码):2011·30·2(200-201)
摘要:

预防X线管损坏对最有效使用X线管、延长X线管寿命有重要作用。本文通过对如何预防X线管阴极损坏、
阳极损坏的各种方法进行分析,得出以下经验:在实际操作X线设备中,必须正确应用X线管的负载曲线
与生热冷却曲线;在维修或操作X线设备中,X线管灯丝不可过热,且避免不必要曝光,同时严格遵守规
则,正确操作;每次曝光后,应有足够的冷却时间;不得碰撞X线管管套。

Effectively preventing the X-ray tube from damage plays an important role for
X-ray tube to be effective use and extend its life.Through analyzing the ways of how to
prevent the damages from X-ray tube cathode and anode,we obtained the following as
advises.In actual operation of X-ray equipment,the operator should correctly use the load
curve and heating-cooling curve.In the maintenance and operation of X-ray
equipment,operator should pay attention not to overheating the X-ray tube filament,avoid it
from unnecessary exposure,and regulate specifications and operate properly.As soon as the
X-ray tube filament complete exposure,operator is better to persist the gap as cooling
time.And the X-ray tube sets do not be crashed all the time.

参考文献:

[1]付建国,裴作升,黄泉荣,等.医学影像设备学[M].北京:高等教育出版社,2005:326-355.
[2]裴作升.医用X线机故障检修方法与技巧[J].北京生物医学工程,2005,24(6):468-469.
[3]裴作升,刘秀珍,龚跃华,等.怎样检修医用X线机[M].武汉:华中科技大学出版社,2006:26-45.
[4]裴作升.F78-Ⅲ型X线机低压电路故障检修一例[J].北京生物医学工程,2006,24(3):336.
[5]姚晓晖.CT机的保养与维护〔J〕.医疗卫生装备,2005,26(7):80.