一种基于压缩传感理论的CT剂量降低方法

A Compressed Sensing-Based CT Dose Reduction Method

作者: 罗海  王淑影  潘文宇  周荷琴 
单位:中国科学技术大学自动化系(合肥230027)
关键词: 计算机断层成像技术;剂量;压缩传感;非均匀采样   
分类号:
出版年·卷·期(页码):2011·30·5(441-446)
摘要:

目的 X射线计算机断层成像技术(computed tomography,CT)辐射剂量高于其它X射线检测。因为剂量累积效应会导致较高的致癌风险,因此降低剂量成为当前CT技术发展过程中亟待解决的问题。方法 降低剂量的手段主要有降低管电流和增加探测器排数,但受相关因素制约,进一步降低剂量面临较大困难。本文提出一种基于X射线衰减水平的自适应非均匀采样方式,通过减少在高管电流区域的投影次数来降低剂量,并引入基于压缩传感的POCS-TVM算法从欠采样数据中恢复CT图像。结果 实验结果表明该方法能在保证图像质量的同时有效降低剂量。结论 基于压缩传感理论的CT剂量降低方法可以在保证重建图像满足临床诊断需求的前提下,有效降低辐射剂量,有利于受检者。

ObjectiveBecause the dose of CT(computed tomography) is higher than other X-ray modalities and the accumulative effect of dose results in high carcinogenic risk,dose reduction is a big challenge in the development of CT technology.MethodsCurrently,methods including decrease of tube current and increase of detector rows have not been efficient enough for further dose reduction.In this paper,we proposed a non-uniform self-adapting sampling method based on the X-ray attenuation.By decreasing the projection views in regions which have high tube current,the dose can be reduced.And POCS-TVM algorithm based on compressed sensing is applied to reconstructing CT image with insufficient sampling data.ResultsThe results of experiments show that this approach can reduce the radioactive dose significantly while ensure the acceptable quality of image.ConclusionsThe CT Dose reduction method based on a compressed sensing can reduce the dose and guarantee the clinical diagnostic quality of the reconstructed CT image,which benefits the subjects.

参考文献:

[1]Hsieh J. Computed Tomography,Principles,Design,Artifacts,and Recent Advances[M].
Bellingham: SPIE,2003.
[2]Tack D,Gevenois PA. Radiation Dose from Adult and Pediatric Multidetector Computed
Tomography[M]. Germany: Springer,2006.
[3]Sodickson A,Baeyens PF,Andriole KP,et al. Recurrent CT,Cumulative Radiation
Exposure,and Associated Radiation-induced Cancer Risks from CT of Adults[J]. Medical
Physics, 2009,251(1):175-184.
[4]Thorsten MB. Computed Tomography:From Photon Statistics to Modern Cone-Beam CT[M].
Germany: Springer,2007.
[5]Gies M,Kalender WA,Wolf H,et al. Dose reduction in CT by anatomically adapted tube
current modulation. I. Simulation studies[J]. Med Phys,1999,26(11):2235-2247.
[6]Sidky EY,Kao CM,Pan X. Accurate image reconstruction from few-views and limited-angle
data in divergent-beam CT[J].  J X-Ray Sci Tech, 2006,14:119-139.
[7]Tang Jie,Nett BE,Chen GH. Performance comparison between total variation (TV)-based
compressed sensing and statistical iterative reconstruction algorithms[J]. Phys Med Biol,
2009,54:5781-5804.
[8]Mannudeep KK,Michael MM,Thomas LT et.al. Techniques and Applications of Automatic Tube
Current Modulation for CT[J]. Radiology, 2004,233:649-657.
[9]Candès E,Romberg J, Tao T. Robust uncertainty principles: Exact signal reconstruct-
tion from highly incomplete frequency informa-tion[J]. IEEE Trans Inform Theory, 2006,52
(2): 489-509.
[10]Yu HY,Ye YB,Zhao SY,et al. Local ROI reconstruction via generalized FBP and BPF
algorithms along more flexible curves[J]. Int J Biomed Imag,2006,Article ID:14989.
[11]庄天戈. CT原理与算法[M]. 上海:上海交通大学出版社,1992.

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com