设为首页 |  加入收藏
首页首页 期刊简介 消息通知 编委会 电子期刊 投稿须知 广告合作 联系我们
一种可变系数的对数脑血管减影算法研究

Logarithmic subtraction algorithm of variable coefficient on cerebral blood vessels

作者: 欧阳丽蓉  邓振生  严昂 
单位:中南大学地球科学与信息物理学院(长沙 410083)
关键词: 数字减影血管造影;可变系数;对数变换 
分类号:
出版年·卷·期(页码):2012·31·1(36-39)
摘要:

目的 针对数字减影血管造影(digital subtraction angiography,DSA)血管图像中非血管结构难以完全去除,严重干扰对血管分析与诊断的问题,本文寻求一种适用于脑血管DSA图像的减影方法。方法 根据X射线在人体组织中的吸收特性,提出一种可变系数的对数减影算法。该算法首先对造影前后的图像分别进行对数变换再相减,然后设计一个自适应的扩大系数,将对数减影后的图像进行灰度动态范围的扩大。结果 对源自临床的一个实际造影图像(蒙片和盈片)分别采用直接减影、一般的对数减影和本文的对数减影算法进行处理,三种方法中本文算法获得的图像非血管结构较少,对比度强。结论 实验结果证明,可变系数对数减影算法能有效消除或降低背景噪声,获得血管灰度一致、对比度较强的血管图像。

Objective  Since the non-vascular structures can hardly be fully expelled from the blood vessels in the image of digital subtraction angiography(DSA),which seriously interferes the analysis and diagnosis of blood vessels,this study looks for a subtraction method suitable for cerebral blood vessels from DSA. Methods According to the absorbing features of X-ray in human bodies,a logarithmic subtraction algorithm of variable coefficient was proposed. Logarithmic transformation was carried on the images of before- and after- angiography respectively,which were then subtracted. Then,a self-adaptive expanded coefficient was used to expand the range of the gray level of the subtraction image. Results Direct subtraction,ordinary logarithmic subtraction and the proposed method were performed on the clinical angiography images respectively in this paper. It showed that the proposed method could obtain images with less non-vascular structures and better contrast than the other two methods. Conclusions The logarithmic subtraction algorithm of variable coefficient might effectively eliminate or decrease background noises,which was helpful to obtain blood vessel images with uniform gray levels and good contrast.

参考文献:

[1]Brody WR. Hybrid Subtraction for Improved Arteriography [J]. Radiology,1981,141(3): 828-831.
[2]Meijering EHW, Niessen WJ,Viergever MA. Retrospective motion correction in digital subtraction angiography: a review [J]. IEEE Trans on Medical Imaging,1999,18(1):2-21.
[3]江林华,陈仕东,柴振明. 基于对数非线性映射的X光数字减影图像的获取与处理[J]. 中国图像图形学报,1999,14(9):773-776.
[4]王群,陈小松,崔至明. 图像增强技术在DSA减影中的应用[J]. 计算机应用与软件,2007,24(10):48-49,69.
[5]师为礼,倪虹. 基于小波变换的数字减影图像配准算法[J].长春理工大学学报,2008,31(4):106-108,115.
[6]刘斌. 脑血管数字减影图像非刚性配准方法研究[D]. 大连:大连理工大学,2007.
[7]齐玲燕,王俊. 基于互信息与边缘梯度相结合的多模医学图像配准方法[J].北京生物医学工程,2011,30(4):359-362.