传声媒质对双界面回波法测量精度的影响

Influence of sound bearing medium on the measurement precision of dual-boundary echo comparison

作者: 张彦  翁善来  郑政 
单位:上海理工大学医疗器械与食品学院(上海 200093)
关键词: 传声媒介;双界面回波法;超声衰减系数;仿组织模块;测量精度 
分类号:
出版年·卷·期(页码):2012·31·1(40-43)
摘要:

目的 双界面回波法(dual-boundary echo comparison,DBEC)是一种有望实现血液超声衰减系数(ultrasonic attenuation coefficient,USAC)无损测量的新方法。使用该方法需要预先测试探头的校正系数,该值反映了声束形状。若将在水中获得的校正系数应用到非水媒质中会带来测量误差,本文通过实验研究这种误差的范围、影响因素以及减小误差的方法。方法 制作折射率与人体组织相当的仿组织超声模块,将其插入声通路中取代一部分水的声程,测量此时的校正系数并与水中所测得的数据进行比较。利用DBEC法对仿组织模块插入前后蓖麻油的USAC进行测量并比较结果。结果 探头通频带内,校正系数误差和蓖麻油USAC的测试误差均在7%以内,但校正系数误差与仿组织模块厚度的关系很复杂。结论 实验结果提示应在仿组织材料中预先测试探头的校正系数,以避免体内测量时DBEC法产生误差。

Objective Dual-boundary echo comparison(DBEC)is a new method of measuring the blood ultrasonic attenuation coefficient(USAC)noninvasively. The correction coefficient of the transducer which reflects the beam shape needs to be determined before the DBEC is applied. The application of the correction coefficient in non-water medium which is obtained in water produces error. This paper studies the range and the influence factors of the error,the methods of minimizing the error. Methods The refractive index of agar phantom is equivalent to human tissues. We insert the agar phantom between the probe and the sample tube,then measure the correction coefficients and compare with those obtained in water. We also measure the USAC of castor oil before and after inserting the tissue-mimicking phantoms and compare the results. Results The experiments show that the maximum error of correction coefficient and the USAC of castor oil are both below 7% within the probe’s frequency bandwidth,however the relationship between the errors and the thickness of the inserted phantom is complicated. Conclusions It is suggest that correction coefficient should be measured in tissue-mimicking phantoms to minimize the error in DBEC.

参考文献:

[1]董彦武,孙永臣,王树恩,等.超声在人的正常与病变组织中的衰减[J].声学学报,1988,13(2):88-93.
[2]Shung KK. Methods for measuring speed,attenuation,absorption,and scattering Diagnostic ultrasound:imaging and blood flow measurements [M]. New York:Taylor & Francis Group,2006:194-195.
[3]冯若. 超声诊断设备原理与设计[M]. 北京:中国医药科技出版社,1992:131-156.
[4]王芳芳,郑政,赵智慧. 一种测量液体声衰减的新方法:双界面回波法[J]. 北京生物医学工程,2010,29(3):249-255.
[5]冯若.超声手册[M]. 南京:南京大学出版社,1999:1076.
[6]Browne J E, Ramnarine K V, Watson A J,et al. Assessment of the acoustic properties of common tissue-mimicking test phantoms[J]. Ultrasound in Medicine & Biology,2003,7(29):1053-1060.

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com