Objective In minimally invasive electrosurgery of sealing blood vessel,it is very important to accurately control the output power of the high-frequency generator. It is a useful method to adjust the output power of high-frequency generator according to the change of blood vessel impedance. In order to obtain the target impedance in the blood vessel sealing,the output voltage and current of the blood vessel circuit are needed to be detected. This article develops an impedance detection system based on LabVIEW. Methods The detection system consists of an isolating and conditioning hardware circuit,data acquisition and processing system based on DAQ card and LabVIEW. Certain non-inductance power resistors with known impedances are measured by this system. Results The experimental results show that this detection system detect the output target impedance with high accuracy. Conclusions This system can be applied in the power control module of high-frequency electrotome used in minimally invasive surgery.
|