设为首页 |  加入收藏
首页首页 期刊简介 消息通知 编委会 电子期刊 投稿须知 广告合作 联系我们
超声多普勒血流和管壁搏动信号的分离研究

Separation method of the Doppler ultrasound blood flow from vessel wall pulsation signals

作者: 朱易顺  彭虎 
单位:中国科学技术大学电子科学与技术系(合肥 230027)
关键词: 超声多普勒;血管壁搏动;血流信号;分离;自适应滤波 
分类号:
出版年·卷·期(页码):2012·31·2(159-163)
摘要:

目的 针对超声多普勒血流检测中,传统的高通滤波法在滤除管壁搏动信号的同时也会滤除低频血流信号的问题,本研究提出一种以心电信号(electrocardiography,ECG)作为参考信号的自适应滤波的方法消除管壁干扰。方法 包括两方面:其一,采用心电信号作为参考信号对超声多普勒信号进行自适应滤波;其二,采用多级自适应滤波并选择不同的参考信号的滤波方案。分别使用上述方法和高通滤波法对仿真的超声多普勒信号进行处理,并将结果进行比较。结果 与传统的高通滤波法相比,该方法在有效抑制管壁搏动信号的同时保留一部分低频血流信号成分。结论 该方法能较准确地提取出完整的血流超声多普勒信号,具有一定的临床应用价值。

Objective In Doppler ultrasound blood flow detection,the traditional high-pass filter would eliminate the low-frequency blood flow signals when it eliminated the vessel wall pulsation signals,so we proposed a method of adaptive filtering which used electrocardiography(ECG)as reference signal to eliminate wall interference. Methods First we utilized the ECG as reference signal for the Doppler ultrasound signals in adaptive filter. Then we adopted the multi-level adaptive filters and selected different reference signals in the filtering programs. Finally we used both methods,the proposed method and high-pass filter,to treat the simulation signals of Doppler ultrasound,and compared the two results. Results The proposed method effectively restrained the wall pulsation signals,and kept a portion of the low blood flow signal components. Conclusions This method could extract the complete Doppler ultrasound blood flow signals accurately,which is valuable for clinical application.

参考文献:

[1]Evans DH,Mcdicken WN. Doppler ultrasound: Physics,instrumentation and signal processing[M].Chichester:John Wiley and Sons,2000.
[2]金大伟,汪源源,王威琪. 空间选择性降噪法提取血流超声多普勒信号[J]. 声学学报,2007,32(3):245-249.
Jin DW,Wang YY, Wang WQ. Extracting Doppler ultrasound blood signals using the spatially selective noise filtration algorithm[J]. Acta Acustica,2007,32(3):245-249.
[3]周彦婷,汪源源. 改进的经验模态分解法分离超声多普勒血流与管壁信号[J]. 声学学报,2010,35(5):495-501.
Zhou YT, Wang YY. Separating Doppler ultrasound blood flow and vessel wall signals by improved empirical mode decomposition methods[J]. Acta Acustica,2010,35(5):495-501.
[4]Wang Y,fish PJ. Comparison of Doppler signal analysis techniques for velocity waveform,turbulence and vortex measurement: a simulation study[J]. Ultrasound in Medicine and Biology,1996,22(5):635-649.
[5]Fish PJ,Bastos CAC,Vaz F. Blood and Wall Signal Simulator for doppler ultrasound signal analysis algorithm development[J].2001 Proceedings of the 23rd Annual EMBS International Conference:135-138.
[6]Diniz PSR. 自适应滤波算法及实现[M]. 第2版.北京:电子工业出版社,2004:48-50.
Diniz PSR. Adaptive Filtering: Algorithms and Practical Implementation [M]. 2nd ed. Liu YL, Jing XJ, Tan GB, translatition. Beijing:Publishing House of Electronics Industry,2004:48-50.
[7]Hoeks APG,van de Vorst JJW,Dabekaussen A,et al. An efficient algorithm to remove low frequency doppler signals in digital Doppler systems[J]. Ultrasound Imaging,1991,13(2):135-144.

服务与反馈: