CT-SD16探测器辐射剂量测量原理和参数设定

Principal analysis on radiation dosimetry of CT-SD16 detector and its parameter setting

作者: 郑海亮  李兴东  刘小丽  冯树理  何宝新  段丛  漆毅   
单位:首都医科大学附属北京世纪坛医院(北京 100038)
关键词: 多排CT;切片探测器;辐射剂量;Barracuda 
分类号:
出版年·卷·期(页码):2012·31·3(278-282)
摘要:

目的 介绍了一种用半导体电离室测量多排CT辐射剂量的原理和方法,为完善我国现有医疗CT检测规范提供技术信息。方法 通过介绍X射线辐射剂量计Barracuda的测量原理和多排CT辐射剂量测量的参数设置依据,并结合在Philip Brilliance iCT 256机器上的应用实例,阐述了用半导体电离室的探测器测量多排CT辐射剂量的关键技术。结果 基于CT扫描参数合理设置探测器参数后,多排CT辐射剂量空间分布情况得以较好展示。100 mm长杆空气电离室的测量范围远小于多排CT线束覆盖范围,需要对多排CT辐射剂量测量方法和技术进行更新。结论 新型探测器有利于揭示多排CT辐射剂量的特性规律。

Objective A mechanism and method of multi-detector or multi-slice CT dosimetry based on semi-conductor slice ionization chamber was introduced to provide technical information for the existing medical CT inspection criterion in China. Methods Through the introduction of X-ray dosimeter Barracuda’s detection principle, the basis of relevant parameter setting, and the application examples of Philip Brillinace iCT 256, the key technologies of multi-slice CT dosimetry was expounded. Results The spatial distribution of CT radiation dosage could be illustrated well by the detector with the slice ionization chamber after proper parameter setting. A pencil ionization chamber with a length of 100 mm was not sufficient to cover the irradiating range by X-ray beam of multi-slice CT, which was expected to renew the detection technology of radiation dosage for multi-slice CT. Conclusions This new detector is helpful to reveal the characteristics of the radiation dosage regularity in multi-slice CT.

参考文献:

[1]JJG 1026-2007, 医用诊断计算机断层摄影装置(CT)X射线辐射源检定规程[S]. 北京: 中国计量出版社, 2007.
[2]白玫, 郑钧正. 多排(层)螺旋CT的辐射剂量表达及其影响因素探讨[J]. 辐射防护, 2008 (1): 1-12.
Bai Mei, Zheng Junzheng. Radiation dosimetry and its influencing factors for the multi-detector/slice spiral CT[J]. Radiation Protection, 2008 (1): 1-12.
[3]徐桓, 赵庆军. 固体切片探测器CT-SD16在CT剂量测量中的应用与研究[J]. 中国医学装备, 2010 (2):17-20.
Xu Heng, Zhao Qingjun. Application and research of solid-state detectors in CT dose measurement[J]. China Medical Equipment, 2010 (2): 17-20.
[4]Bauhs JA Vrieze TJ, Primak AN, Bruesewitz MR, McCollough CH. CT dosimetry: comparison of measurement techniques and devices[J]. Radiographics, 2008 (28): 9.
[5]Mori S, Nishizawa K, Ohno M,et al. Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scanner[J]. Br J Radiol, 2006, 79(947): 888-892.

服务与反馈:
文章下载】【加入收藏
提示:您还未登录,请登录!点此登录
 
友情链接  
地址:北京安定门外安贞医院内北京生物医学工程编辑部
电话:010-64456508  传真:010-64456661
电子邮箱:llbl910219@126.com