改良型CT模拟机定位精度分析

Analysis of positioning accuracy for improved CT simulator

作者:               陈定科  黄灿辉  董勇  梁耀文  陈伟洪          
单位:           东莞市石龙人民医院肿瘤科(广东东莞523326)    
关键词:           放射治疗;CT模拟机;定位      
分类号:
出版年·卷·期(页码):2013·32·4(418-420)
摘要:

目的 研究改良型CT模拟机的定位精度,探讨其是否满足精确放射治疗的要求。方法 在头颈部固体模体内逐一随机预埋一直径为3cm金属圆球10次,将此金属圆球假设为靶区,然后分别用改良型CT模拟机定位并在直线加速器下复位,依次得出X(左右方向)、Y(前后方向)、Z(头脚方向)方向的10次定位误差。结果 10次定位误差在X、Y、Z方向的误差分别为(0.13±0.23)mm、(0.26±0.36)mm和(0.45±0.43)mm。结论 改良型CT模拟机定位精度满足精确放射治疗小于3mm的定位精度需要。

Objective To study the positioning accuracy of improved CT simulator and to investigate whether it reaches the precision requirement of radiation therapy. Methods A metal ball of 3cm diameter was embedded in the solid phantoms of head and neck 10 times at random. Assuming the metal ball was the target area,we obtained 10 positioning errors in X-axis (left to right),Y-axis (front to back) and Z-axis (head to foot) in turn when improved CT simulator was applied for positioning and linear accelerator was applied for resetting. Results The means and standard errors of the 10 positioning errors in X-axis,Y-axis,Z-axis,were (0.13±0.23) mm,(0.26±0.36) mm,(0.45±0.43) mm,respectively. Conclusions The positioning precision of the improved CT simulator reaches the precision requirement of radiation therapy —within 3mm.

参考文献:

[1]邓小武,黄劭敏,钟宁山,等.调强放射治疗的物理剂量验证[J].癌症,2001,20(10):1092-1094.

Deng Xiaowu,Huang Shaomin,Zhong Ningshan,et al. Dosimetry verification and quality assurance test of IMRT[J].Chinese Journal of Cancer,2001,20(10):1092-1094.

[2]何正中,戴建荣,乐文友.调强放疗中摆位误差对脊髓剂量的影响[J].中华放射学肿瘤杂志,2007,16(2):132-135.

He Zhengzhong,Dai Jianrong,Yue Wenyou. Dose uncertainty in the spinal cord from setup errors in intensity-modulated radiation therapy[J]. Chinese Journal of Radiation Oncology,2007,16(2):132-135.

[3]林承光,李国文,刘晓臻,等.鼻咽癌调强放射治疗摆位不确定度对危及器官计划体积(PRV)影响的研究[J].癌症,2008,27(3):327-330.

Lin Chengguang,Li Guowen,Liu Xiaozhen,et al. Influence of patient setup error on planning risk volumes for nasopharyngeal carcinoma patients receiving intensity-modulated radiation therapy[J]. Chinese Journal of Radiation Oncology,2008,27(3):327-330.

[4]李文杰,林承光,陈立新.CT扫描/重建参数对三维治疗计划系统影像的影响[J].肿瘤学杂志,2005,11(5):323-326.

Li Wenjie,Lin Chengguang,Chen Lixin. Influence of CT scan & reconstruction parameters on imaging in 3-dimensional treatment planning system[J]. Journal of Chinese Oncology,2005,11(5):323-326.