The static mechanical and fatigue performance testing method of implants of orthopaedic?nickel titanium shape memory alloy embracing compressive connector
Objective Implants of orthopaedic?Nickel titanium shape memory alloy embracing compressive connector has important application value in orthopedics because of its unique shape memory effect and excellent mechanical properties. However
its static mechanical and fatigue properties in practical clinical applications still need to be systematically evaluated. For this reason
this study establish a set of scientific and standardized static mechanical and fatigue performance test methods
which can provide a reliable basis for the evaluation of the performance of the nickel?titanium shape memory alloy bone grafting ring hugger. Methods First
based on the clinical use characteristics and mechanical performance requirements of Nickel titanium shape memory alloy embracing compressive connector
a static mechanical and fatigue performance evaluation system was constructed
and a corresponding standard test method was developed. Secondly
we chose the rib embracing compressive connector
femur embracing compressive connector and patella embracing compressive connector which had been registered and listed in the market as the test objects
and carried out the static mechanical test (5 pieces in each group) and the fatigue test ( 6 pieces in each group)
respectively. The testing process simulated the implanted human body and the usage environment (including installation methods and body temperature conditions) to obtain key mechanical parameters such as bending structural stiffness
bending strength
bending rigidity
number of fatigue life cycles
maximum bending moment and loading load. Finally
the feasibility and repeatability of the test method were verified by bias analysis. Results AccordingThe test results showed that the established static mechanical and fatigue performance test method could effectively obtain the key mechanical parameters of Nickel titanium shape memory alloy embracing compressive connector. The deviation analysis showed that the test data had high scientific and repeatability
which verified that the test method was feasible and repeatable
and the data was scientific and valid. Conclusions In this study
a set of perfect static mechanical and fatigue performance evaluation methods for Implants of orthopaedic?Nickel titanium shape memory alloy embracing compressive connector has been established to provide a scientific basis for the evaluation of safety and effectiveness in clinical implantation applications
to provide a theoretical basis for the subsequent design optimization and clinical validation
and is of great significance in promoting the performance enhancement of orthopaedic implantable devices.