Application of artificial intelligence?assisted compressed sensing technique on 3D T1?weighted imaging of intracranial vessel walls
|更新时间:2026-01-26
|
Application of artificial intelligence?assisted compressed sensing technique on 3D T1?weighted imaging of intracranial vessel walls
Vol. 44, Issue 5, (2025)
作者机构:
作者简介:
基金信息:
DOI:
CLC:
Published:2025
稿件说明:
移动端阅览
张月青, 赵元宾, 张瑞, et al. Application of artificial intelligence?assisted compressed sensing technique on 3D T1?weighted imaging of intracranial vessel walls[J]. 2025, 44(5).
DOI:
张月青, 赵元宾, 张瑞, et al. Application of artificial intelligence?assisted compressed sensing technique on 3D T1?weighted imaging of intracranial vessel walls[J]. 2025, 44(5).DOI:
Application of artificial intelligence?assisted compressed sensing technique on 3D T1?weighted imaging of intracranial vessel walls
ACS)技术加速颅内血管壁高分辨率三维 T1 加权黑血成像( three⁃dimensional T1⁃weighted black blood imaging
3D T1_HRVWI)的可行性
并与临床常规采用的并行采集技术( parallel imaging
PI)进行对比分析。 方法 前瞻性纳入 47 例脑血管疾病患者
行 ACS 加速(实验组)和 PI 加速(对照组)的 3D T1_HRVWI 扫描。 客观比较 ACS 组和 PI 组的序列扫描时长、图像中血管壁的信噪比以及相对于血管腔的对比噪声比
主观评估两组图像总体质量、颅内血管显示情况及病灶诊断价值
并进行统计分析。 结果 相较于 PI3D T1_HRVWI
ACS 3D T1_HRVWI 的扫描时长缩短 43%;ACS 3D T1_HRVWI 图像中颈内动脉、基底动脉、椎动脉、大脑前动脉、大脑中动脉、大脑后动脉血管壁的信噪比均显著升高(P<0 001);两组图像的总体质量评分和颅内血管的显示评分均未见显著差异(P>0 05);两组图像的病灶诊断价值相当(P>0 05)。 结论 ACS 技术实现了 3D T1_HRVWI 序列更快的采集速度
同时保持了较好的成像质量
为开发兼顾扫描效率与诊断精度的脑血管影像扫描方案提供了循证依据
在脑血管疾病的精准诊断中展现出重要价值。
Abstract
Objective To explore the feasibility of artificial intelligent?assisted compressed sensing (ACS) technology in accelerating high?resolution three?dimensional T1?weighted black blood imaging (3D T1_HRVWI) of intracranial blood vessel walls
in comparison with the parallel imaging (PI) technique. Methods 47 patients with cerebrovascular diseases were prospectively enrolled in this study. ACS accelerated and PI accelerated 3D T1_HRVWI scanning were both performed. The scanning duration
signal?to?noise ratio (SNR) of vessel walls in 3D T1_HRVWI images and the contrast noise ratio (CNR) of vessel walls relative to vascular lumens were objectively compared between the ACS group and the PI group. Also
the overall quality of 3D T1 _HRVWI images
intracranial artery visualization and diagnostic value of lesions were subjectively evaluated. Statistical analysis was performed. Results Compared with PI 3D T1_HRVWI
the scanning time of ACS 3D T1 _HRVWI was shortened by 43%. In ACS 3D T1_HRVWI images
the SNR of internal carotid artery
basilar artery
vertebral artery
anterior cerebral artery
middle cerebral artery and posterior cerebral artery were significantly higher ( P < 0?? 001 ). There were no significant differences on scores for overall quality and intracranial artery visualization betweenthe two groups (P>0?? 05). The diagnostic value of the two sequences was equal (P>0?? 05). Conclusions Compared with PI
ACS can shorten the acquisition time of 3D T1_HRVWI while ensuring the image quality
which provides an evidence?based basis for developing neurovascular imaging scanning protocols and shows important value in the accurate diagnosis of neurovascular diseases.