Objective To evaluate the detection function of the new domestically developed online calibration device for ventilators developed by National Institute of Metrology
China
and provide experimental evidence for its application in metrology testing. Methods This study used the SV600 invasive ventilator produced by Shenzhen Mindray Company as the tested device. The detection performance of the online calibration device developed by National Institute of Metrology
China was evaluated by using four universal ventilator detection devices:the handheld VT900A produced by FLUKE in the United States
the desktop PF300 and handheld H3 produced by IMTMEDICAL in Switzerland
and the handheld SINO?VT?1 produced by Beijing Bangrui Company. The detection parameters included tidal volume
respiratory frequency
peak airway pressure
positive end expiratory pressure
and oxygen concentration. Results The local online calibration device met all requirements of the national metrological standard testing specifications. There were statistically significant differences (P<0.05) in the accuracy of tidal volume detection and the repeatability of oxygen concentration detection within certain ranges compared to other ventilator testing equipment. Conclusions Preliminary evaluation shows that under the same testing conditions
the new domestically produced ventilator online calibration device meets the technical requirements in detecting respiratory rate
peak airway pressure
and positive end expiratory pressure parameters
and has significant advantages in the error value of tidal volume and the detection accuracy of oxygen concentration. It can be put into clinical use for further verification.