Application of quality control circle in improving the management quality of medical equipment in the resuscitation room
Objective Through the quality control circle activity
to reduce the failure rate of ECG monitor caused by human factors and improve the management quality of medical equipment in Emergency Room. Methods Firstly
the professional quality control circle of Medical Engineering Department and Emergency Room was established. Then
the failure of the ECG monitor in Emergency Room from September 2021 to September 2022 was collected and analyzed with the help of management tools such as brainstorming
Gantt chart and fishbone chart
and the PDCA cycle was developed and implemented according to the failure reasons. Finally
the failure rate of ECG monitor before and after the implementation of countermeasures was compared
and the effectiveness of implementation countermeasures was evaluated by chi?square test. Results The average repair report times of ECG monitor in emergency room decreased from 3.25 times/ month to 1 times/ month
and the human failure rate of ECG monitor decreased from 43.33% to 15.38% ( P < 0.01 )
the difference was statistically significant. The target achievement rate of this quality control circle activity was 126%
and the improvement rate was 64.50%. Conclusions The introduction of quality control circle into the fault analysis of ECG monitor in emergency room can effectively reduce the failure rate of ECG monitor caused by human factors
avoid medical risks and save medical costs
improve clinical efficiency and patient satisfaction
and provide a new idea and method for the use and management of medical equipment.