LYU Bin,YANG Huifang.Application of hydrogen peroxide disinfection technology in prevention and control of infection of medical devices[J].Beijing Biomedical Engineering,2026,45(3):285-290.
LYU Bin,YANG Huifang.Application of hydrogen peroxide disinfection technology in prevention and control of infection of medical devices[J].Beijing Biomedical Engineering,2026,45(3):285-290.DOI: 10.3969/j.issn.1002-3208.2026.03.009.
Application of hydrogen peroxide disinfection technology in prevention and control of infection of medical devices
Because there are many kinds of medical devices and different requirements and methods of infection prevention and control, the qualified rate of disinfection and cleaning quality is low. Therefore, this paper introduces hydrogen peroxide disinfection technology to study the application effect of this technology in the prevention and control of infection of medical devices.
Methods
2
This study selected 9 895 reusable medical devices which were representative in a specific clinical environment of a hospital as research objects. These medical devices were widely used in various clinical departments of the hospital, involving the diagnosis, treatment and nursing process of a variety of diseases. On the basis of detailed analysis of various risk factors of infection, the infection risk coefficient corresponding to each risk factor was calculated accurately by scientific statistical method and risk assessment model. Through the quantitative analysis and comparison of these coefficients, six evaluation items with higher risk of infection were finally selected. According to the characteristics and infection risk degree of different medical devices, hydrogen peroxide disinfectant, application equipment and application method were selected, and the application effect of disinfection technology in the prevention and control of infection of medical devices was analyzed in combination with the qualified rate of disinfection of devices, the incidence rate of hospital infection and the sterilization effect of devices.
Results
2
After the application of disinfection technology in infection prevention and control of medical devices, the qualified rate of disinfection and cleaning quality of medical devices was more than 80%, which effectively improved the effect of infection prevention and control management of medical devices. Adenosinetriphosphate (ATP) content was always maintained within 200 amol/ piece, had a good disinfection effect, and the biological indicators were qualified.
Conclusions
2
This technology can effectively improve the implementation effect of infection prevention and control measures for medical devices, not only meet the high quality requirements of clinical disinfection and cleaning of medical devices, but also provide a feasible and effective solution for optimizing infection prevention and control measures for medical devices.
Xiao P,Tang YL. Effect of ultrasonic cleaning with multi⁃enzyme cleaner combined with low temperature plasma sterilization on washing passing rate and sterilization effect of medical apparatus and instruments[J]. Chinese Nursing Research,2023,37(10):1878-1880.
Qian YQ, Ni TT, Fan JJ, et al. Study on the sterilization effect of complex chlorine disinfectant for medical devices on flexible endoscopes[J]. Chinese Journal of Disinfection,2023,40(4):246-249,253.
Suenaga Y , Kawano H , Takamatsu T ,et al.Ultrasonic-combined plasma bubbling for adherent bacteria disinfection on medical equipment[J].Plasma Chemistry and Plasma Processing, 2022,42(3):575-586.
Zhao N, Bai YT, Wang H, et al. Application value of medical quality management in the management of automatic washer disinfector[J]. China Medical Equipment,2023,20(3):164-168.
Zheng A , Zhang C .The application value of sanitary wipes and medical alcohol+gauze disinfection method in the management and use of medical equipment[J].Journal of Clinical and Nursing Research, 2023, 7(6):162-166.
Guo QY, Liu YL. Impact of PDCA management in sterile supply centres on the quality of external medical device cleaning[J]. Nursing Practice and Research,2023,20(9):1406-1410.
Liao HB, Li YY, Hou XL, et al. Influencing factors of unqualified cleaning of medicaldevices in central sterile supply[J]. Henan Medical Research,2023,32(7):1278-1282.
Yan YZ, Chen XX, Liu J, et al. Effect of work revamping based on root cause analysis on disinfection supply center[J]. Journal of Clinical Nursing in Practice,2022,8(11):154-156.
Ultrasound image segmenation of thyroid nodules based on multi-scale feature extraction and multi-feature fusion
Acoustic analysis of swallowing sounds for patients with swallowing disorders prospects for the application of evaluation and screening
Exploring the molecular mechanisms of lung adenocarcinoma and construction of a prognostic risk model based on bioinformatics methods
Construction and practice of a dual-track “scientific research + science popularization” collaborative extension activity system for the medical physics course under the context of new medicine
Design and implementation of the MRI equipment monitoring and early warning system