纸质出版:2025
移动端阅览
闫鹏伟, 缪祥文, 申光, 等. 骨科植入器械镍钛形状记忆合金接骨环抱器的静态力学与疲劳性能测试方法[J]. 北京生物医学工程, 2025,44(5).
闫鹏伟, 缪祥文, 申光, et al. The static mechanical and fatigue performance testing method of implants of orthopaedic?nickel titanium shape memory alloy embracing compressive connector[J]. 2025, 44(5).
目的 骨科植入器械镍钛形状记忆合金接骨环抱器因其独特的形状记忆效应和优异的力学性能
在骨科领域具有重要应用价值。 然而
其在实际临床应用中的静态力学与疲劳性能仍需系统评估。 为此
本研究建立一套科学、规范的静态力学与疲劳性能测试方法
为镍钛形状记忆合金接骨环抱器的性能评价提供可靠依据。 方法 首先
基于镍钛形状记忆合金接骨环抱器的临床使用特点及力学性能要求
构建静态力学与疲劳性能评价体系
并制定相应的标准测试方法。 其次
选择已注册上市的肋骨接骨环抱器、股骨接骨环抱器及髌骨接骨环抱器作为测试对象
分别进行静态力学试验(每组 5 件)和疲劳试验(每组 6 件)。 测试过程中
模拟植入人体和使用环境(包括安装方式及体温条件)
获取弯曲结构刚度、弯曲强度、弯曲刚度、疲劳寿命循环次数、最大弯矩及加载载荷等关键力学参数。 最后
通过偏差分析验证测试方法的可行性与重复性。 结果 所建立的静态力学与疲劳性能测试方法能够有效获取镍钛形状记忆合金接骨环抱器的关键力学参数。 偏差分析显示
测试数据具有较高的科学性与重复性
验证了测试方法具有可行性且重复性较好
数据科学有效。 结论 建立了一套完善的骨科植入器械镍钛形状记忆合金接骨环抱器静态力学和疲劳性能评价方法
为其在临床植入应用中的安全性和有效性评价提供科学依据
为后续的设计优化和临床验证提供理论基础
对推动骨科植入器械的性能提升具有重要意义。
Objective Implants of orthopaedic?Nickel titanium shape memory alloy embracing compressive connector has important application value in orthopedics because of its unique shape memory effect and excellent mechanical properties. However
its static mechanical and fatigue properties in practical clinical applications still need to be systematically evaluated. For this reason
this study establish a set of scientific and standardized static mechanical and fatigue performance test methods
which can provide a reliable basis for the evaluation of the performance of the nickel?titanium shape memory alloy bone grafting ring hugger. Methods First
based on the clinical use characteristics and mechanical performance requirements of Nickel titanium shape memory alloy embracing compressive connector
a static mechanical and fatigue performance evaluation system was constructed
and a corresponding standard test method was developed. Secondly
we chose the rib embracing compressive connector
femur embracing compressive connector and patella embracing compressive connector which had been registered and listed in the market as the test objects
and carried out the static mechanical test (5 pieces in each group) and the fatigue test ( 6 pieces in each group)
respectively. The testing process simulated the implanted human body and the usage environment (including installation methods and body temperature conditions) to obtain key mechanical parameters such as bending structural stiffness
bending strength
bending rigidity
number of fatigue life cycles
maximum bending moment and loading load. Finally
the feasibility and repeatability of the test method were verified by bias analysis. Results AccordingThe test results showed that the established static mechanical and fatigue performance test method could effectively obtain the key mechanical parameters of Nickel titanium shape memory alloy embracing compressive connector. The deviation analysis showed that the test data had high scientific and repeatability
which verified that the test method was feasible and repeatable
and the data was scientific and valid. Conclusions In this study
a set of perfect static mechanical and fatigue performance evaluation methods for Implants of orthopaedic?Nickel titanium shape memory alloy embracing compressive connector has been established to provide a scientific basis for the evaluation of safety and effectiveness in clinical implantation applications
to provide a theoretical basis for the subsequent design optimization and clinical validation
and is of great significance in promoting the performance enhancement of orthopaedic implantable devices.