目的 针对运动模糊图像复原中
传统频谱法存在的检测误差大、抗噪性弱及有效检测范围有限等问题
本文提出一种融合高斯拉普拉斯(Laplacian of Gaussian
LoG)滤波与 Radon 变换的边缘增强频谱分析法
旨在实现点扩散函数(point spread function
PSF) 参数的精准估计。 方法 基于脑部MRI 仿真运动模糊模型
分析频谱中明暗条纹的分布特征
采用 LoG 滤波提取频谱亮条纹边缘
抑制中心宽条纹与 Gibbs 现象导致的干扰
生成保留方向特征的离散边缘点集;利用 Radon 变换包容非共线点集的抗噪特性(离散边缘点沿角度 θ 投影时
真实边缘贡献相干叠加
噪声点投影随机抵消)
显著提升峰信噪比
进而精准定位模糊角度并计算模糊长度;采用配对 t 检验
对传统中心亮条纹检测方法与本文方法的 PSF 参数估计误差进行统计学分析比较。 结果 本研究方法的角度估计平均误差 0.08°
显著低于传统方法的 3.28°
长度估计平均误差 0.15 像素
传统方法为 0.88 像素
有效角度检测范围由传统方法的±60°扩展到 0~ 180°
且组间误差差异均达极显著水平(P<0.001)。 结论 本方法通过 LoG 滤波与 Radon 变换的协同机制
避免了对中心条纹完整性的依赖
解决了宽条纹导致的检测失效问题
同时有效抑制了噪声和 Gibbs 现象导致的干扰
显著提高了运动模糊 PSF 参数估计的精度与鲁棒性
为医学影像运动伪影消除提供可靠的技术基础。
Objective Traditional spectrum analysis methods for motion blur image restoration suffer from significant detection errors
weak noise resistance
and a limited effective detection range. To address these limitations
this study proposes an edge?enhanced spectrum analysis method integrating Laplacian of Gaussian (LoG) filtering and Radon transform
aiming to achieve accurate estimation of point spread function ( PSF) parameters. Methods Based on a simulated motion blur model for brain MRI
the distribution characteristics of bright and dark stripes in the spectrum were analyzed. LoG filtering was employed to extract the edges of the bright spectral stripes
suppressing interference from the central wide stripe and Gibbs phenomenon
thereby generating a discrete set of edge points preserving directional features. Utilizing the noise?resistant property of Radon transform concerning non?collinear point sets ( where projections of genuine edge points coherently superimposed along angle θ
while noise projections cancelled out randomly)
the peak signal?to?noise ratio was significantly enhanced. This facilitated precise localization of the blur angle and calculation of the blur length. A paired t?test was conducted for statistical comparison of the PSF parameter estimation errors between the traditional central bright stripe detection method and the proposed method. Results The average angle estimation error of the proposed method was 0. 08°
significantly lower than the 3.28° of the traditional method. The average length estimation error was 0.15 pixels
compared to 0.88 pixels for the traditional method. The effective angle detection range was extended from ±60° for the traditional method to 0 to 180°. The differences in errors between the groups were statistically highly significant ( P < 0.001 ). Conclusions The proposed method
through the synergistic mechanism of LoG filtering and Radon transform
avoids reliance on the integrity of the central stripe and resolves the detection failure issue caused by wide stripes. It effectively suppresses interference from noise and the Gibbs phenomenon
significantly enhancing the accuracy and robustness of motion blur PSF parameter estimation. This provides a reliable technical foundation for eliminating motion artifacts in medical imaging.